2.5 Oxygen Concentration is High
Chapter 2: Troubleshooting Procedures
Page 2 - 58 Isolette® Infant Incubator (Model C2000) Service Manual
2.5 Oxygen Concentration is High
1. Perform the following:
a. If the High Oxygen % alarm sounds, press the Silence/Reset
softkey (A) to silence it for 4 minutes (see figure 2-21 on page
2-58).
Figure 2-21. Hood/Shell Assembly
b. Go to RAP 2.31 on page 2-94.
The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
2. Perform the following:
a. Slowly tilt the hood assembly (B) back until it comes to rest.
b. Remove the foam mattress (C), the mattress tray (D), and the x-
ray tray (E) from the shell assembly (F) (see figure 2-22 on page
2-59).