EasyManuals Logo

Dräger Isolette C2000 Service Manual

Dräger Isolette C2000
524 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #255 background imageLoading...
Page #255 background image
4.20 Heater Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 54 Isolette® Infant Incubator (Model C2000) Service Manual
Replacement
1. Connect the heater assembly (D) to the AC wiring assembly (H),
and route the cables down into the shell assembly (G).
CAUTION:
Always replace Nylok®
1
screws; do not reuse them. Equipment
damage could occur.
2. Install the three new Nylok® screws (F) to secure the heater
assembly (D) to the shell assembly (G).
3. Screw the heater radiator (E) onto the heater assembly (D).
4. Install the heater/impeller cover (C) on the shell assembly (G).
5. Install the main deck (B) on the shell assembly (G).
6. Slide the two mattress tilt bars (A) through their openings in the main
deck (B).
7. Install the mattress, the mattress tray, and the x-ray tray on the unit
(refer to procedure 4.7 on page 4-20).
8. Carefully lower the hood assembly to its closed position.
9. To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.
1. Nylok® is a registered trademark of Nylok Fastener Corporation.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Isolette C2000 and is the answer not in the manual?

Dräger Isolette C2000 Specifications

General IconGeneral
BrandDräger
ModelIsolette C2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals