2.42 Skin 1 Probe Does Not Work Properly
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 109
2
2.42 Skin 1 Probe Does Not Work Properly
1. Perform the following:
a. If the Skin 1 Probe Fail alarm sounds, press the Silence/Reset
softkey (A) to silence it for 5 minutes.
Figure 2-37. Skin 1 Probe
b. Check the skin probe (B).
The skin probe (B) is securely attached to the infant’s skin.
Yes No
↓→Securely attach the skin probe (B) to the infant’s skin. If
this solves the problem, go to “Final Actions” on page 2-24.
Otherwise, go to step 5.
2. Disconnect the skin probe (B) from the 1 skin connector at the
sensor module assembly (C). The Skin Probe Disconnect alarm
sounds.
Yes No
↓→Go to step 5.
3. The unit is in Air Mode. Perform the following: