2.42 Skin 1 Probe Does Not Work Properly
Chapter 2: Troubleshooting Procedures
Page 2 - 110 Isolette® Infant Incubator (Model C2000) Service Manual
a. Press the Silence/Reset softkey (A) to silence the Skin Probe
Disconnect alarm.
b. Set the unit in Skin Mode.
This solves the problem.
Yes No
↓→Go to step 6.
4. Go to “Final Actions” on page 2-24.
5. Check the mode of operation. The unit is in Skin Mode.
Yes No
↓→Set the unit in Skin Mode. If this solves the problem, go to
“Final Actions” on page 2-24. Otherwise, go to step 6.
6. The Connect Skin 1 Probe alarm sounds.
Yes No
↓→Go to RAP 2.38 on page 2-104. If this solves the problem,
go to “Final Actions” on page 2-24. Otherwise, go to step 7.
7. Perform the following:
a. Connect the skin probe (B) to the 1 skin connector on the sensor
module assembly (C).
b. Access the System Test (1 of 3) menu (see “Diagnostic Menus”
on page 2-35). Skin Probe 1 reads as Connected.
Yes No
↓→Securely connect the skin probe (B) to the sensor module
assembly (C). If this solves the problem, go to “Final
Actions” on page 2-24. Otherwise, go to step 8.
8. Access the Data Summary (1 of 2) screen (see “Diagnostic Menus”
on page 2-35). The Skin 1 Temp reads 0.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
9. Replace the skin probe (B) (refer to procedure 4.1 on page 4-5). The
problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
10.Replace the sensor module assembly (C) (refer to procedure 4.2 on
page 4-6). This solves the problem.