Specifications
Chapter 1: Introduction
Page 1 - 14 Isolette® Infant Incubator (Model C2000) Service Manual
Oxygen Control Description
For Isolette® Infant Incubator oxygen control specifications, see table
1-3 on page 1-14.
Table 1-3. Oxygen Control Specifications
Humidity Description
For Isolette® Infant Incubator humidity specifications, see table 1-4 on
page 1-14.
Table 1-4. Humidity Specifications
Feature Dimension
Oxygen display range 18% to 100%
Oxygen display resolution 1% oxygen
Oxygen display accuracy (100%
calibration)
± 3%
Oxygen display accuracy (21% calibra-
tion)
± 5%
Oxygen set resolution 1%
Oxygen setpoint range 21% to 65%
Feature Dimension
Operating range 5% to 99% relative humidity
(RH), non-condensing
Storage range 0% to 99% RH, non-con-
densing
Operating time without refilling > 24 hours maximum at 85%
RH
Humidity control accuracy at 10% to
80% at 68 F (20 C) to 104 F (60 C)
± 6% RH
Maximum humidity level at the incuba-
tor’s set temperature of 102 F (39 C)
> 85% RH
Humidity setpoint range 30% to 95% RH

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Isolette C2000 and is the answer not in the manual?

Dräger Isolette C2000 Specifications

General IconGeneral
BrandDräger
ModelIsolette C2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals