2.10 Controller Failure #5—Display Test Failure
Chapter 2: Troubleshooting Procedures
Page 2 - 66 Isolette® Infant Incubator (Model C2000) Service Manual
2.10 Controller Failure #5—Display Test Failure
1. Turn the controller assembly off, and then turn it back on. The
screen flashes briefly, and then turns black.
Yes No
↓→Replace the electroluminescent (EL) display (refer to
procedure 4.15 on page 4-43). If this solves the problem,
go to “Final Actions” on page 2-24. Otherwise, go to step 2.
2. Replace the EL central processing unit (CPU) P.C. board assembly
(refer to procedure 4.15 on page 4-43). This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
3. Go to “Final Actions” on page 2-24.