2.41 Power Failure Alarm Sounds
Chapter 2: Troubleshooting Procedures
Page 2 - 108 Isolette® Infant Incubator (Model C2000) Service Manual
2.41 Power Failure Alarm Sounds
1. Perform the following:
a. Silence the Power Failure alarm for 15 minutes by pressing the
Silence/Reset softkey.
b. Plug the unit into an appropriate power source.
This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
2. Go to “Final Actions” on page 2-24.