2.40 Humidity System Is Not Working Properly
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 107
2
d. Using a calibrated hygrometer, monitor the relative humidity level
within the hood assembly (D).
Within 30 to 60 minutes, the hygrometer and Humidity display do
not both read 50% ± 9%.
Yes No
↓→Go to “Final Actions” on page 2-24.
3. For assistance, call Technical Support at (800) 437-2437.