EasyManua.ls Logo

Dräger Isolette C2000 - Humidity System Is Not Working Properly

Dräger Isolette C2000
524 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.40 Humidity System Is Not Working Properly
Chapter 2: Troubleshooting Procedures
Page 2 - 106 Isolette® Infant Incubator (Model C2000) Service Manual
2.40 Humidity System Is Not Working Properly
1. Perform the following:
a. If the Low Humidity alarm sounds, press the Silence/Reset key
(A) to silence it for 5 minutes (see figure 2-36 on page 2-106).
Figure 2-36. Humidity System
b. Withdraw the humidity tray (B) to the fill position.
c. Check the humidity reservoir (C).
The humidity reservoir (C) is filled.
Yes No
↓→Fill the humidity reservoir (C) with distilled water (refer to
procedure 6.7 on page 6-34). If this solves the problem, go
to “Final Actions” on page 2-24. Otherwise, go to step 2.
2. Perform the following:
a. Warm the incubator to 95 °F (35 °C).
b. At Display 1, press the Humidity softkey to turn on the humidity.
c. Set the humidity setpoint to 50%.

Table of Contents

Related product manuals