7.1 Weighing Scale
Chapter 7: Accessories
Isolette® Infant Incubator (Model C2000) Service Manual Page 7 - 7
7
6. Connect the scale platform cable assembly (H) to the Weight
connector on the sensor module assembly (I).
7. Ensure that there is sufficient slack in the scale platform cable
assembly (H) between the edge of the hood assembly (A) and the
scale platform (G):
a. Raise the hood assembly (A) to its full open position.
b. Fully withdraw the mattress tray (F) from the hood assembly (A).
WARNING:
Ensure all sensor leads are properly routed. Use cable management
clips to avoid entanglement and possible infant injury.
8. Using the cable clips (J) provided, secure the scale platform cable
assembly (H) to the inside wall of the hood assembly (A).
9. Calibrate the weighing scale (refer to procedure 6.4 on page 6-25).
10.To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.
Removal
Perform the installation procedure in reverse order.