Specifications
Chapter 1: Introduction
Isolette® Infant Incubator (Model C2000) Service Manual Page 1 - 13
1
Electrical Description
For Isolette® Infant Incubator electrical specifications, see table 1-2 on
page 1-13.
Table 1-2. Electrical Specifications
Feature Dimension
Power requirements of an incubator
with a fixed height pedestal stand
(100V and 120V models)
100V/120V ± 10%, 50/60 Hz,
630 W maximum
Power requirements of an incubator
with a variable height adjustable (VHA)
pedestal stand (100V and 120V mod-
els)
100V/120V ± 10%, 50/60 Hz,
1900 W maximum
Power requirements of an incubator
with a fixed height or a VHA pedestal
stand (220V and 240V models)
220V or 240V ± 10%,
50/60 Hz, 2376 W maximum
Convenience outlets (100V and 120V
models)
100V/120V ± 10%, 50/60 Hz,
500 W maximum
Convenience outlets (220V and 240V
models)
220V or 240V, 500 W
maximum
Chassis current leakage (110V and
120V models)
300 μA or less
Chassis current leakage (220V and
240V models)
500 μA or less