Page 1 - 12 Isolette® Infant Incubator (Model C2000) Service Manual
Air Mode temperature range in
Temperature Override Mode
98.6 °F (37.0 °C) to 102.2 °F
(39.0 °C)
Skin Mode temperature range 93.2 °F (34.0 °C) to 98.6 °F
(37.0 °C)
Skin Mode temperature range in
Temperature Override Mode
93.2 °F (34.0 °C) to 100.4 °F
(38.0 °C)
Temperature rise time, typical at 72 F
(22 C) ambient
< 35 minutes
Temperature variability < 0.5 C maximum
Temperature overshoot < 0.5 C maximum
Temperature uniformity < 0.8 °C
Correlation of the indicated air temper-
ature to the actual incubator tempera-
ture after temperature equilibrium is
reached
0.8 °C