Lubrication Requirements
Chapter 6: General Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 6 - 9
6
Lubrication Requirements
WARNING:
Follow the product manufacturer’s instructions. Failure to do so could
result in personal injury or equipment damage.
CAUTION:
Do not use silicone-based lubricants. Equipment damage could occur.
Oilite®
1
bearings and bushings are utilized in several places on the
system. By retaining oil, the pores give a self-lubricating quality to the
bearings and bushings. If any silicone-based lubricant is applied to the
bearings and bushings or anywhere else on the system, this self-
lubricating quality is neutralized.
It is safe to apply the following lubricants to the system (see table 6-1 on
page 6-9):
Table 6-1. Lubricants
1. Oilite® is a registered trademark of Beemer Precision, Incorporated.
Part Number Description
8252 (100) 2 oz m-1 oil (apply to Oilite® bearings and bush-
ings)
SA3351 (100) 4 oz lithium grease