4.34 Access Grommet
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 96 Isolette® Infant Incubator (Model C2000) Service Manual
4.34 Access Grommet
Tools required: None
Removal
1. On a unit with a Series 00 hood assembly (A), slide the blue slide
access panel latches, and open the access panel (B) (see figure 4-
17 on page 4-38).
or
On a unit with a hood assembly (A) later than Series 00, rotate the
access panel pawl latch knobs (C), and open the access panel (B).
Figure 4-50. Access Grommet
2. Pivot the access panel (B) to the full-open position so it hangs
straight down.
3. Remove any wiring or cables from the access grommet (D).
4. Remove the access grommet (D) from the hood assembly (A).