2.22 Sensor Module Is Out of Position
Chapter 2: Troubleshooting Procedures
Page 2 - 80 Isolette® Infant Incubator (Model C2000) Service Manual
2.22 Sensor Module Is Out of Position
1. The sensor module assembly (A) is properly positioned in the hood
or calibration fixture (see figure 2-26 on page 2-80).
Figure 2-26. Sensor Module Assembly
Yes No
↓→Properly position the sensor module assembly (A) in the
hood or calibration fixture. If this solves the problem, go to
“Final Actions” on page 2-24. Otherwise, go to step 2.
2. Replace the sensor module assembly (A) (refer to procedure 4.2 on
page 4-6). This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
3. Go to “Final Actions” on page 2-24.