2.21 Communication Between the Sensor Module and the Controller Fails
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 79
2
2. Replace the sensor module-to-controller cable assembly (D) (refer
to procedure 4.23 on page 4-65). The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
3. Replace the sensor module assembly (refer to procedure 4.2 on
page 4-6). This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
4. Go to “Final Actions” on page 2-24.