4.30 Heat Shield
Chapter 4: Removal, Replacement, and Adjustment Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 4 - 83
4
Figure 4-41. Heat Shield
4. Release the heat shield (F) from the access panel (C) by pressing
on the two heat shield latches (G).
Replacement
1. Perform the removal procedure in reverse order.
2. To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.