6.7 Filling the Humidity Reservoir
Chapter 6: General Procedures
Page 6 - 34 Isolette® Infant Incubator (Model C2000) Service Manual
6.7 Filling the Humidity Reservoir
Tools required: Distilled water
1. Release the latch (A) that secures the humidity tray (B) in the shell
assembly (C), and withdraw the humidity reservoir (D) until it stops
at the fill position (see figure 6-10 on page 6-34).
Figure 6-10. Humidity Reservoir
CAUTION:
To avoid equipment damage, use only distilled or sterile distilled water.
Sterile water alone is not an acceptable substitute for distilled water.
2. Fill the humidity reservoir (D) with distilled water.
3. To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.