6.6 Disassembly and Assembly for Cleaning
Chapter 6: General Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 6 - 33
6
d. Slide the humidity tray (N) and the humidity reservoir (I) into the
unit.
12.Inspect the tubing access grommets, and perform one of the
following:
If a tubing access grommet is distorted or torn, replace it (refer to
procedure 4.34 on page 4-96).
If a tubing access grommet is not distorted or torn, install it on the
hood assembly (refer to procedure 4.34 on page 4-96).
13.Install an access door gasket on each access door (refer to
procedure 4.27 on page 4-76).
14.If the incubator is to be gas sterilized, sterilize it.
15.Install a new access door cuff onto each access door gasket (refer
to procedure 4.26 on page 4-74).
16.Inspect the air intake microfilter, and perform one of the following:
If the air intake microfilter is discolored or dirty, replace it (refer to
procedure 4.19 on page 4-51).
Install the air intake microfilter and the air intake microfilter cover
onto the unit (refer to procedure 4.19 on page 4-51).
17.To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Isolette C2000 and is the answer not in the manual?

Dräger Isolette C2000 Specifications

General IconGeneral
BrandDräger
ModelIsolette C2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals