2.45 Sensor Module Assembly Fails
Chapter 2: Troubleshooting Procedures
Page 2 - 114 Isolette® Infant Incubator (Model C2000) Service Manual
2.45 Sensor Module Assembly Fails
1. The sensor module is properly installed on the hood assembly.
Yes No
↓→Go to RAP 2.22 on page 2-80. If this solves the problem,
go to “Final Actions” on page 2-24. Otherwise, go to step 2.
2. The sensor module is properly connected to the unit.
Yes No
↓→Go to RAP 2.21 on page 2-78. If this solves the problem,
go to “Final Actions” on page 2-24. Otherwise, go to step 3.
3. Replace the sensor module assembly (refer to procedure 4.2 on
page 4-6). This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
4. Go to “Final Actions” on page 2-24.