2.25 Access Panel Assembly Does Not Latch Properly
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 83
2
2.25 Access Panel Assembly Does Not Latch Properly
1. Check the area around the access panel assembly (A) for
obstruction (see figure 2-27 on page 2-83). The area around the
access panel assembly (A) is free from obstruction.
Figure 2-27. Side Access Panel
Yes No
↓→Remove the obstruction from the access panel assembly
(A). If this solves the problem, go to “Final Actions” on
page 2-24. Otherwise, go to step 2.
2. The access panel assembly (A) is a Series 00 model.
Yes No
↓→Go to step 5.
3. Slide the blue slide of the access panel latch assemblies (B) back
and forth. The blue slides of the access panel latch assemblies (B)
slide properly.
Yes No
↓→Replace the access panel assembly (A) (refer to
procedure 4.10 on page 4-29). If this solves the problem,

Table of Contents