2.25 Access Panel Assembly Does Not Latch Properly
Chapter 2: Troubleshooting Procedures
Page 2 - 84 Isolette® Infant Incubator (Model C2000) Service Manual
go to “Final Actions” on page 2-24. Otherwise, call
Technical Support at (800) 437-2437 for assistance.
4. Go to step 6.
5. Rotate the access panel knob and latch assemblies (C). The access
panel knob and latch assemblies (C) rotate properly.
Yes No
↓→Replace the access panel knob and latch assembly (C)
(refer to procedure 4.24 on page 4-70). If this solves the
problem, go to “Final Actions” on page 2-24. Otherwise, go
to step 6.
6. Replace the access panel assembly (A) (refer to procedure 4.10 on
page 4-29). This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
7. Go to “Final Actions” on page 2-24.