2.4 Oxygen Concentration is Low
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 57
2
3. Check the sleeves on the iris entry ports (D). All sleeves are
properly installed.
Yes No
↓→Properly install the sleeves on the iris entry ports (D) (refer
to procedure 4.11 on page 4-33). If this solves the
problem, go to “Final Actions” on page 2-24. Otherwise, go
to step 4.
4. Check the access grommets (E). The access grommets (E) are
properly installed.
Yes No
↓→Properly install the access grommets (E) (refer to
procedure 4.34 on page 4-96). If this solves the problem,
go to “Final Actions” on page 2-24. Otherwise, go to step 5.
5. Check the air intake microfilter cover (F). The air intake microfilter
cover (F) is properly secured.
Yes No
↓→Tighten the two thumbscrews (G) that secure the air intake
microfilter cover (F) to the unit. If this solves the problem,
go to “Final Actions” on page 2-24. Otherwise, go to step 6.
6. Perform the following:
a. Loosen the two thumbscrews (G) on the air intake microfilter
cover (F).
b. Remove the air intake microfilter cover (F) from the unit.
c. Check the air intake microfilter (H).
The air intake microfilter (H) is installed.
Yes No
↓→Install the air intake microfilter (H) (refer to procedure 4.19
on page 4-51). If this solves the problem, go to “Final
Actions” on page 2-24. Otherwise, go to step 7.
7. Go to RAP 2.31 on page 2-94. This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
8. Go to “Final Actions” on page 2-24.