2.47 Scale Is Not Working Properly
Chapter 2: Troubleshooting Procedures
Page 2 - 118 Isolette® Infant Incubator (Model C2000) Service Manual
7. Place a weight of known value less than 15 lb (7 kg) on the mattress
(D). The correct weight appears, and when the Weight Sample bar
is filled, a beep sounds and the weight is locked into the Weight
display.
Yes No
↓→Go to step 10.
8. Perform the following:
a. Press the Home key to return to Display 2.
b. Press the Weigh softkey.
The display again displays the value of the weight on the mattress
(D).
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
9. The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
10.Check the connection of the scale assembly cable (E) at the sensor
module assembly (F). The scale assembly cable (E) is securely
connected to the sensor module assembly (F).
Yes No
↓→Securely connect the scale assembly cable (E) to the
sensor module assembly (F). If this solves the problem, go
to “Final Actions” on page 2-24. Otherwise, go to step 11.
11. Replace the sensor module assembly (F) (refer to procedure 4.2 on
page 4-6). The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
12.Replace the scale assembly (refer to procedure 7.1 on page 7-5).
This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
13.Go to “Final Actions” on page 2-24.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Isolette C2000 and is the answer not in the manual?

Dräger Isolette C2000 Specifications

General IconGeneral
BrandDräger
ModelIsolette C2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals