2.31 Air Intake Is Not Working Properly
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 95
2
WARNING:
A dirty air intake microfilter may affect oxygen concentrations and/or
cause carbon dioxide build-up. Check the filter routinely, and change it
at least every 3 months or when it is visibly dirty. Failure to do so could
result in infant injury.
2. Inspect the air intake microfilter (C). The air intake microfilter (C) has
no visible dirt, is dry, and has been used for less than three months.
Yes No
↓→Replace the air intake microfilter (C) (refer to procedure
4.19 on page 4-51). If this solves the problem, go to “Final
Actions” on page 2-24. Otherwise, call Technical Support
at (800) 437-2437 for assistance.
3. Go to “Final Actions” on page 2-24.