2.3 Skin Set Temperature Cannot Be Achieved or Maintained
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 55
2
Yes No
↓→Place the unit in Skin Mode. If this solves the problem, go
to “Final Actions” on page 2-24. Otherwise, go to step 4.
4. Check the skin probe (D). The skin probe (D) is properly secured to
the infant’s skin.
Yes No
↓→Properly secure the skin probe (D) to the infant’s skin. If
this solves the problem, go to “Final Actions” on page 2-24.
Otherwise, call Technical Support at (800) 437-2437 for
assistance.
5. Call Technical Support at (800) 437-2437 for assistance.