2.18 Heater Thermocouple Does Not Work Properly
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 75
2
2. Replace the AC wiring harness assembly (D) (refer to procedure
4.22 on page 4-60). The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
3. Replace the heater assembly (refer to procedure 4.20 on page
4-52). This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
4. Go to “Final Actions” on page 2-24.