2.34 Incubator Takes Longer Than One Hour to Warm Up to the Air Set Temperature
Chapter 2: Troubleshooting Procedures
Page 2 - 100 Isolette® Infant Incubator (Model C2000) Service Manual
Yes No
↓→Replace the access grommet (C) (refer to procedure 4.34
on page 4-96). If this solves the problem, go to “Final
Actions” on page 2-24. Otherwise, go to step 5.
5. Go to RAP 2.18 on page 2-74. This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
6. Go to “Final Actions” on page 2-24.