2.32 X-ray Tray Does Not Operate Properly
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 97
2
Yes No
↓→Remove the obstruction from the x-ray tray (A). If this
solves the problem, go to “Final Actions” on page 2-24.
Otherwise, go to step 3.
3. Replace the x-ray tray (A) (refer to procedure 4.7 on page 4-20).
The problem still exists.
Yes No
↓→Go to “Final Actions” on page 2-24.
4. Replace the mattress tray (B) (refer to procedure 4.7 on page 4-20).
This solves the problem.
Yes No
↓→For assistance, call Technical Support at (800) 437-2437.
5. Go to “Final Actions” on page 2-24.