Safety Tips
Chapter 1: Introduction
Page 1 - 20 Isolette® Infant Incubator (Model C2000) Service Manual
WARNING:
For the 21% oxygen calibration, the unit must be equipped with a
standard sensor module lock. Failure to use a standard sensor module
could result in incorrect calibration. Infant injury could occur.
WARNING:
For the 100% oxygen calibration procedure, the incubator hood must
be equipped with the 100% calibration fixture. Failure to use the 100%
calibration fixture could result in incorrect calibration. Infant injury could
occur.
WARNING:
Oxygen flow rates cannot be used as an accurate indication of oxygen
concentration in an incubator. Continuously monitor the oxygen
concentrations with a calibrated oxygen analyzer. Failure to do so could
result in personal injury or equipment damage.
WARNING:
Ensure all sensor leads are properly routed. Use cable management
clips to avoid entanglement and possible infant injury.
SHOCK HAZARD:
Ensure that the building power source is compatible with the electrical
specifications shown on the right side of the incubator and the variable
height adjustable pedestal stand. For proper grounding reliability,
connect the power cord only to a properly marked, three-wire, hospital-
grade or hospital-use receptacle. Do not use extension cords.
SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.
SHOCK HAZARD:
Unplug the power cord from the controller assembly. Failure to do so
could result in personal injury or equipment damage.