2.29 Mattress Tray Does Not Tilt Properly
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 91
2
Yes No
↓→Replace the right-hand mattress tilt mechanism (refer to
procedure 4.32 on page 4-90). If this solves the problem,
go to “Final Actions” on page 2-24. Otherwise, go to step 5.
4. Rotate the right-hand mattress tilt mechanism knob (B) clockwise
until it stops. The mattress tray (A) is level.
Yes No
↓→Replace the right-hand mattress tilt mechanism (refer to
procedure 4.32 on page 4-90). If this solves the problem,
go to “Final Actions” on page 2-24. Otherwise, go to step 5.
5. Rotate the left-hand mattress tilt mechanism knob (B) clockwise.
The left end of the mattress tray (A) is at a 12 angle.
Yes No
↓→Replace the left-hand mattress tilt mechanism (refer to
procedure 4.32 on page 4-90). If this solves the problem,
go to “Final Actions” on page 2-24. Otherwise, call
Technical Support at (800) 437-2437 for assistance.
6. Rotate the left-hand mattress tilt mechanism knob (B)
counterclockwise. The mattress tray (A) is level.
Yes No
↓→Replace the left-hand mattress tilt mechanism (refer to
procedure 4.32 on page 4-90). If this solves the problem,
go to “Final Actions” on page 2-24. Otherwise, call
Technical Support at (800) 437-2437 for assistance.
7. Go to “Final Actions” on page 2-24.