4.9 Motor Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 28 Isolette® Infant Incubator (Model C2000) Service Manual
7. If the unit is equipped with a humidity system, install it (refer to
procedure 7.3 on page 7-15).
8. Install the impeller assembly on the unit (refer to procedure 4.8 on
page 4-22).
9. Install the mattress, the mattress tray, and the x-ray tray on the unit
(refer to procedure 4.7 on page 4-20).
10.Install the hood assembly on the unit (refer to procedure 4.5 on page
4-14).
11. Install the controller assembly in the shell assembly (refer to
procedure 4.4 on page 4-11).
12.Plug the power cord into the controller assembly (A).
13.To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.