4.10 Access Panel
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 30 Isolette® Infant Incubator (Model C2000) Service Manual
4. Pull the two mattress tilt bars (E) up through the main deck (F), and
remove them from the unit (see figure 4-12 on page 4-30).
Figure 4-12. Mattress Platform Assembly
5. Close and latch the access panel (C) (see figure 4-11 on page
4-29).
6. Using a 1/8" hex head wrench, remove the two shoulder screws (G)
and the two flat washers (H) that secure the access panel hinges (I)
to the hood assembly (A) (see figure 4-13 on page 4-31). Retain the