4.26 Access Door Cuff
Chapter 4: Removal, Replacement, and Adjustment Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 4 - 75
4
2. Ensure proper installation of the access door cuff (C) by checking
the following:
The access door cuff (C) has a small opening at its center.
The access door (B) latches with slight pressure and opens when
the access door latch (A) is pressed.
3. To ensure proper operation of the Isolette® Infant Incubator,
perform the “Function Checks” on page 2-6.