6.1 Installation and Set-up
Chapter 6: General Procedures
Page 6 - 14 Isolette® Infant Incubator (Model C2000) Service Manual
WARNING:
The hood/shell assembly must attach to the pedestal/stand using the
screws provided. Failure to do so could result in the hood/shell
assembly separating from the pedestal stand when sufficiently tilted,
particularly with the hood open. Personal injury or equipment damage
could occur.
2. Install the four ¼"-20 x 1¾" Nylok®
1
screws (D) to secure the
hood/shell assembly (A) on the pedestal stand (B).
1. Nylok® is a registered trademark of Nylok Fastener Corporation.