6.6 Disassembly and Assembly for Cleaning
Chapter 6: General Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 6 - 29
6
3. Slowly tilt the hood assembly (B) back until it is fully open.
4. Slide the x-ray tray (C) out of the mattress tray (D) (see figure 6-8 on
page 6-30).
5. If the incubator is equipped with a weighing scale, remove it (refer to
procedure 7.1 on page 7-5).
6. Remove the mattress (E) and the mattress tray (D) (refer to
procedure 4.7 on page 4-20).
7. Pull the mattress tilt bars (F) up through the main deck (G).
8. Remove the main deck (G).
9. Remove the heater/impeller cover (H).
WARNING:
Allow 45 minutes for the heater assembly to cool. Failure to do so could
result in personal injury.
10.Switch off the unit, and allow 45 minutes for the heater assembly to
cool.
11. Remove the heater radiator and the heater assembly from the unit
(refer to procedure 4.20 on page 4-52).
12.Remove the impeller assembly (refer to procedure 4.8 on page
4-22).