System Failure Messages
Chapter 2: Troubleshooting Procedures
Page 2 - 34 Isolette® Infant Incubator (Model C2000) Service Manual
Heater Failed 2 The heater thermocouple wires
are open or shorted.
Go to RAP 2.18 on page
2-74.
Humidity Heater
Failure
The humidity heater draws too
much current.
Go to RAP 2.19 on page
2-76.
Motor Failed For software version 2.06, the fan
motor speed exceeds ± 450 rpm;
for previous software versions, the
fan motor speed exceeds ± 200
rpm.
Go to RAP 2.20 on page
2-77.
Sensor Disconnect Communications fail between the
sensor module and the controller.
Go to RAP 2.21 on page
2-78.
Sensor Out of
Position
The sensor module is not in the
hood or the calibration position.
Go to RAP 2.22 on page
2-80.
Stuck Key A stuck key is detected. Go to RAP 2.23 on page
2-81.
Check Settings The non-volatile memory (NVM)
integrity test fails.
Go to RAP 2.24 on page
2-82.