Diagnostic Menus
Chapter 2: Troubleshooting Procedures
Page 2 - 36 Isolette® Infant Incubator (Model C2000) Service Manual
2. After viewing the current settings, perform one of the following:
• Return the System Configuration menu to the original settings as
shown in table 2-9 on page 2-36 by pressing the Hard Default key
(B).
• View the Diagnostic Information menu by pressing the Diag Info
key (C). Go to “Entering the Diagnostic Information Menu” on page
2-37.
• Enter the desired system configuration settings, and press the
Silence/Reset key (A) to return to Display 1.
Table 2-9. System Configuration Factory Default Settings
Feature Setting
Temperature control mode Air mode
Air set temperature 95.0 °F (35.0 °C)
Skin set temperature 97.7 °F (36.5 °C)
Humidity control mode OFF
Humidity set value 50% RH
Oxygen control mode OFF
Oxygen calibration level 21%
Temperature units °C
Weight units Kg
Trend Air, 2 hours
Language English
Altitude < 10000' (3048 m)