Diagnostic Menus
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 41
2
Table 2-11. System Test (1 of 3) Menu
After viewing the current system test results, perform one of the
following:
• Check the system checksum by pressing the Checksum key (G).
• Check the system RAM by pressing the System RAM key (H).
• Check the display RAM by pressing the Display RAM key (I).
• View the System Test (2 of 3) screen by pressing the Display
Selection key (J). Go to “System Test (2 of 3) Menu” on page 2-42.
• Return to the Diagnostic Info menu by pressing the Home key (F).
Go to “Entering the System Configuration Menu” on page 2-35.
Information Item Description
O
2
monitor cell This displays the installation status of the oxygen
monitor fuel cell.
O
2
control cell This displays the installation status of the oxygen
control cell.
Humidity sensor This displays the installation status of the humid-
ity sensor.
Scale This displays the connection status of the scale
assembly.
Skin probe 1 This displays the connection status of the skin
temperature probe 1.
Skin probe 2 This displays the connection status of the skin
temperature probe 2.
Hood door This displays the open/closed status of the front
or back access panel assembly.
Heater This displays the on/off status of the unit’s hea-
ter.
Humidifier This displays the on/off of the humidity evapora-
tor assembly.
RTC battery This displays the pass/fail status of the real-time
clock (RTC) battery.