83
────────────────────────────────────────────────────
5.2 Making Settings
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
NOTE
Range Compression ratio
50 ms/DIV × 1/2
20 ms/DIV × 1/10
10 ms/DIV × 1/20
Time axis range
/DIV
Sampling period
1μs 10μs 100μs 1ms 10 ms 100 ms
10 ms
20 ms
50 ms
100 ms
200 ms
500 ms
1s
2s
5s
10 s
30 s
1 min
2 min
5 min
10 min
30 min
1 hour
Method 2 Screen: STATUS 1, DISPLAY
1. Call up the STATUS 1 or DISPLAY screen.
2. Use the
TIME/DIV key to set the time axis range.
The
TIME/DIV key can be used regardless of where the flashing cursor is
located, if the STATUS 1 or DISPLAY screen is displayed.
The sampling period cannot be changed on the DISPLAY screen.
While the printer always outputs the data at the measurement
magnification in recording mode, the waveform on the screen is reduced in
size at the ratio shown in the table below, depending on the time-axis range.
When the printer is ON, the recording length, "continuous" cannot be set at
a time axis range setting of 10 ms to 200 ms/DIV.
Setting value of the sampling period
: can be set,
: can not be set