7.3 Making the Settings of the 8936 ANALOG UNIT
────────────────────────────────────────────────────
Displayed screen
(zero position: 100%
)
0V
100%
50%
0%
Displayed screen
(zero position: 0%
)
100%
50%
0%
Displayed screen
(zero position: 50%
)
100%
50%
0%
1600
LSB
2047
0
-2048
Memory recorder function
Recorder function
A/D data
Magnification ratio: ×
1
4095
RMS recorder functio
n
Magnification ratio ×1/2 ×1 ×2 ×5 ×10
No. of full-scale
LSBs
3200 1600 800 320 160
Zero position
adjustment area
(Memory,
recorder)
0 to 100 -50 to 150 -150 to 250 -450 to 550 -950 to 1050
Zero position
adjustment area
(RMS recorder)
-50 to 100 -150 to 150 -350 to 250 -950 to 550 -1950 to 1050
NOTE
Reference
The zero positions are shown in the figure below.
It is possible to display the hidden portion of the waveform by setting "0 V"
to a proper percentage on the display.
Available measurement range of each range is approximately +, - 12.5 times
voltage axis range setting. Any part beyond available voltage measurement
range is not loaded as waveform but indicated as out of range.
According to zero position and magnification/compression ratio of voltage
axis, voltage range displayed in waveform display screen varies but
available measurement range does not.