145
────────────────────────────────────────────────────
7.7 Making the Settings of the 8940 F/V UNIT
────────────────────────────────────────────────────
NOTE
Setting the Pull-Up Resistance
Set the pull-up resistance on or off.
Pull-up resistance is used when connecting to an open collector output signal.
For normal measurements, disable the pull-up resistance (set to OFF).
Method Screen: CHANNEL 1, DISPLAY
1. Call up the CHANNEL 1 or DISPLAY screen.
2. Move the flashing cursor to the point shown
in the figure on the left.
3. Use the function keys to make the selection.
: Pull-up resistance is enabled (ON).
(for connection to open collector)
: Pull-up resistance is disabled (OFF).
Pull-up settings cannot be determined from the DISPLAY or CHANNEL 1
screen. Check through the function display that appears when the flashing
cursor is moved to the "input coupling" item.
Pull-up ON or OFF is printed when doing a list print.