157
────────────────────────────────────────────────────
7.9 Making the Settings of the 8947 CHARGE UNIT
────────────────────────────────────────────────────
(Example 2) When sensor sensitivity is 0.05 (pC/(m/s
2
))
Multiply value by 2 and take 0.1 (pC/(m/s
2
)) as sensor sensitivity value as
setting.
Configure scaling.
[Conversion ratio scaling]
EU/V: 0.1/0.05 = 2.0000E+00
Offset: 0.0000E+00
EU: m/s
2
[2-point scaling]
0.0000E+00 0.0000E+00
0.0000E+00 0.0000E+00
EU: m/s
2
Setting the Measurement Range
Set the measurement range.
Sensor sensitivity modifies measurement range. Adjust sensor sensitivity
before setting measurement range (see Section 15.2.7).
Method Screen: CHANNEL 1, DISPLAY
1. Move the flashing cursor to the point shown
in the figure on the left.
2. Use the function keys or the jog control or to
make the selection.
: Move the cursor up in the selection window.
: Move the cursor down in the selection
window.