218
────────────────────────────────────────────────────
9.4 Scaling Function [ SCALING ]
────────────────────────────────────────────────────
NOTE
: Voltage high point SC
: Scaling high point
: Voltage low point SC
: Scaling low point
X
Scaling setting
Variable setting
NOTE
When the 2-point scaling is used, scaling is performed using the following
equation.
The ranges for the parts enclosed in dotted lines are as follows.
-9.9999E+9 value of enclosed part -1.0000E-9
Value of enclosed part = 0
-1.0000E-9 value of enclosed part +9.9999E+9
A setting outside of the above range cannot be made.
Relationship between scaling function and variable function
Convert the measured value to a physical quantity.
Set the unit.
Set the upper and lower limits to be used after conversion
to a physical quantity.
Results are the same regardless of the order in which
and are set.
In
, set the variable function with the value after scaling.
(Set the unit (EU) on the SCALING screen.)
When using both the scaling and variable functions, the upper and lower
limits for the variable display after scaling are as follows.
-9.9999E29 setting value -1.0000E - 29
Setting value = 0
+1.0000E-29 setting value +9.9999E+29
When a setting outside of the above range is attempted, a warning indication
is given.