9.4 Scaling Function [ SCALING ]
────────────────────────────────────────────────────
NOTE
V
H
: Voltage high point SC
H
: Scaling high point
V
L
: Voltage low point SC
L
: Scaling low point
Scaling setting
Variable setting
NOTE
・ When the 2-point scaling is used, scaling is performed using the following
equation.
・The ranges for the parts enclosed in dotted lines are as follows.
-9.9999E+9 ≦ value of enclosed part ≦ -1.0000E-9
Value of enclosed part = 0
-1.0000E-9 ≦ value of enclosed part ≦ +9.9999E+9
A setting outside of the above range cannot be made.
Relationship between scaling function and variable function
・Convert the measured value to a physical quantity.
・Set the unit.
・Set the upper and lower limits to be used after conversion
to a physical quantity.
・Results are the same regardless of the order in which
and are set.
・In
, set the variable function with the value after scaling.
(Set the unit (EU) on the SCALING screen.)
・ When using both the scaling and variable functions, the upper and lower
limits for the variable display after scaling are as follows.
-9.9999E+29 ≦setting value ≦ -1.0000E - 29
Setting value = 0
+1.0000E-29 ≦ setting value ≦ +9.9999E+29
・ When a setting outside of the above range is attempted, a warning indication
is given.