EasyManua.ls Logo

Olympus BondMaster 600 - Figure 6-38 the Metal to Metal Disbonds Application; Figure 6-39 the CAL Screen

Olympus BondMaster 600
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10045-01EN, Rev. E, August 2016
Applications
149
3. Select the Metal To Metal Disbonds application, and then press to accept
(see Figure 6-38 on page 149).
Figure 638 The Metal To Metal Disbonds application
4. If the calibration menu does not appear automatically, press and hold the CAL
NULL key ( ).
5. Hold the probe in the air. The BondMaster 600 should automatically pick the best
operating frequency for the probe. When in doubt, press CAL (C key) or change
the FREQ (D key) using the knob.
6. Press DONE (E key) [see Figure 6-39 on page 149].
Figure 639 The CAL screen

Table of Contents

Other manuals for Olympus BondMaster 600

Related product manuals