EasyManua.ls Logo

Olympus BondMaster 600 - Figure 6-18 the Skin to Core Disbonds (Tapered) Application; Figure 6-19 the Swept Figure between Two Divisions

Olympus BondMaster 600
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10045-01EN, Rev. E, August 2016
Applications
135
Figure 618 The Skin To Core Disbonds (Tapered) application
To calibrate the signals
1. Press the MAIN menu key ( ) once, and then use the knob to set the START
FREQ (C key) to 10 kHz and the STOP FREQ (D key) to 40 kHz.
2. Place the probe tips on a good part of the standard, press the GAIN key ( )
and then use the knob to adjust the gain until the swept figure is contained
between two divisions inside the alarm box (see Figure 6-19 on page 135).
Figure 619 The swept figure between two divisions

Table of Contents

Other manuals for Olympus BondMaster 600

Related product manuals