EasyManua.ls Logo

Olympus BondMaster 600 - Figure 6-46 the Delamination and Laminates Inspection Application; Figure 6-47 the CAL Screen

Olympus BondMaster 600
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10045-01EN, Rev. E, August 2016
Applications
155
3. Select the Delamination and Laminates Inspection application, and then press
to accept (see Figure 6-46 on page 155).
Figure 646 The Delamination and Laminates Inspection application
4. If the calibration menu does not appear automatically, press and hold the CAL
NULL key ( ).
5. Hold the probe in the air. The BondMaster 600 should automatically pick the best
operating frequency for the probe. When in doubt, press CAL (C key) or change
the FREQ (D key) using the knob.
6. Press DONE (E key) [see Figure 6-47 on page 155].
Figure 647 The CAL screen

Table of Contents

Other manuals for Olympus BondMaster 600

Related product manuals