Software Reference 177
V4.2 LabChip GX User Manual PerkinElmer
Peak Table View (Continued)
Fragment End (sec) - (RNA assays only) Amount of time from
injection to the falling edge of the fragment in seconds.
Fragment Start (sec) - (RNA assays only) Amount of time from
injection to the leading edge of the fragment in seconds.
FWHM - The width of the peak (Full Width at Half Maximum),
measured in seconds.
Height - The value at the apex of the peak minus the local baseline
start value.
Instrument Name - The name of the instrument that was used to
run the plate and create the data file.
Molarity (nmol/l) - (DNA and gDNA assays only) Displays the
molarity of the peak.
Migration Time - Center - The time from injection to the peak apex
in seconds.
Peak Index - The index number of the peaks in the sample. Marker
peaks, peaks before the lower marker, and peaks after the upper
marker are not numbered.
Peak Number - The order in which the peaks were detected for
each well. All peaks in each well are numbered, starting at 1.
Plate File - The name of the original plate data file. This name
cannot be changed.
Plate Name - The name of the plate defined in the workspace.
Plate Number - The number of the plate in the same order as the
plates are displayed in the Plate View or Plate List.
RNA Fragment - (RNA assays only) The name of the fragment
based on its size falling within a certain range set in the assay
definition. (These definitions are visible only to assay developers).
Sample Name - The sample name defined in the Sample Name File
selected in the Advanced Tab on the Start Run Window. The
Sample name can also be changed in the Well Table View or the
Gel View.
Signal Noise - The statistical standard deviation of 1 second of
data just before the Peak Find Start Time.