V4.2 LabChip GX User Manual PerkinElmer
Filter View (Continued)
When using multiple filters that include peak properties, if
AND(Peak) is selected, a single peak in the well must meet all
selected filter properties. See the examples below, which each use
two filters:
• Height > 100 AND(Peak) Conc > 50 selects any well that
contains a single peak that meets both criteria.
• Height >100 OR Conc > 50 selects any well that contains a peak
that meets either (or both) criteria.
• Height >100 AND(Peak) Height <50 does not select any wells
because no peak can have a height that is both less than 50 and
greater than 100. Use OR to select wells that have peaks that
match either criteria. Use AND(Well) to select wells that contain
peaks that match both criteria.
Expected Peak Filters can contain multiple size peaks in the same
filter as shown above. Expected Peak filters with multiple peak
sizes in the same filter always select only wells that contain all the
peaks listed.