Commands Listed in Alphabetical Order
ATTENUATION 25
X
sets the P7260 probe to ±3 V dynamic range with 6 GHz
bandwidth and 25X attenuation.
VTERMsource A
UTO
sets the P7380SMA probe voltage termination source to
auto.
VTERMsourc
e INTerna l
sets the P7380SMA probe voltage termination source
to internal.
VTERMsourc
e EXTerna l
sets the P7380SMA probe voltage termination source
to external.
Examples
AUXIN:PROBE:SET "ATTEN 25X" sets the P7260 probe to the selection for
±3 V dynamic range with 3 GHz bandwidth and 25X attenuation.
AUXIN:PROBE:SET? might return :CH2:PROBE: SET "ATTENUATION 5X"
indicating that the P7260 probe is set to the selection for ±0.75 V dynamic range
with 6 GHz bandwidth and 5 X attenuation.
AUXIN:PROBE:SET? "ATTEN" might return :CH2:PROBE:SET
"ATTENUATION 25X"
, indicating that the P7260 probe is set to the selection for
±3 V dynamic range with 3 GHz bandwidth and 25X attenuation.
AUXIn:PRObe:SIGnal
This command sets or queries aspects of probe accessory user interfaces. The
available arguments for this command will vary depending on the accessory you
attach to the instrument.
Group
Vertical
Syntax
AUXIn:PRObe:SIGnal {PASS|BYPass}
AUXIn:PRObe:SIGnal?
Arguments
PASS opens a relay passing your signal to the instrument.
BYPass closes a relay preventing your signal from reaching the instrument.
During probe d egauss, the signal should be bypassed.
Examples
AUXIN:PROBE:SIGNAL PASS opens a relay passing your signal to the instrument.
AUXIN:PROBE:SIGNAL? might return :AUXIN:PROBE:SIGNAL BYPASS,
indicating that the relay is closed prev enting your signal from reaching the
instrument.
2-96 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual